Insti-Fluor Multi Color test slidesInsti-Fluor Test Slides for detecting bacteria using fluorescence microscopy. These slides are pre-loaded with either DAPI, SYTO, or PI. Remove the seal top from all three slides Add 10µl sample to the center of each well Spread the sample over the well surface Dry at 42C for about 5min While the samples are drying set your pipetter to 7µl and find the immersion oil We use Zeiss oil with a n=1.518You may use any oil, just make sure it has NO INTRINSIC FLUORESCENCE Finished drying Add 7µl immersion oil to the center of the wellIt's better to under-estimate the volume than over-fill. Apply the coverslipsBe careful as you have only one try. The coverslips cannot be removed. Gently press the coverslip all around to create a permanent seal.The slide can now be decontaminated as per the usual SOP. These slides are ready to be examined.